  • 麻醉呼吸机.png

Animal intelligent anesthesia ventilator


Anesthesia ventilator specially designed for pets. Suitable for pet hospitals, pet clinics and animal laboratories.

Pre-consultation
Description Parameters Resources

Anesthesia ventilator specially designed for pets. Suitable for pet hospitals, pet clinics and animal laboratories.

Its technical parameters can meet the respiratory needs of cats, dogs and animals within 100 kg. International standard interface, which can be directly matched with other brands of anesthesia machines.


Features

1. 5.7 inch TFT LCD display

2. The shuttle button can quickly set parameters

3. Tidal volume setting: 20ml~1500ml

4. Pneumatic electric control, built-in battery

5. Breathing modes: VCV (Volume Controlled Ventilation), PCV (Pressure Controlled Ventilation), SIGH (Sigh Mode), Standby (Standby Mode), Manual (Manual Mode)

6. Multi-waveform display: pressure-time, flow-time, capacity-time

7. Built-in electronic PEEP valve, equipped with flow sensor module

Ventilation parameter setting range:

1. Tidal volume: 20~1500ml
2. Respiratory rate: 1~100bmp
3. Inspiratory time: 0.1~10.0s
4. Breathing ratio: 4:1~1:10
5. Apnea time: 0~60% of inspiratory time
6. Positive end-expiratory pressure: off, 3~30cmH2O
7. Pressure control level: 5~70cmH2O
Monitoring of ventilation parameters:

1. Inspiratory tidal volume: 0~2500ml
2. Expiratory tidal volume: 0~2500ml
3. Minute ventilation: 0~60L/min
4. Respiratory rate: 0~100bmp
5. Breathing ratio: 9:1~1:99
6. Airway peak pressure: 0~100cmH2O
7. Average airway pressure: 0~100cmH2O
8. Positive end-expiratory pressure: 0~100cmH2O
9. Inhalation platform pressure: 0~100cmH2O
10. Inhaled oxygen concentration: 15~100%
11. Compliance: 0~300ml/cmH2O
12. Airway resistance: 0~600cmH2O (L/S)
Electrical Specifications:
1. Input voltage: one-way AC100~240V
2. Input frequency: 50/60Hz
3. Input power: <60VA
Internal battery:
1. Battery type: NiMH battery
2. Shutdown time: > 10 minutes
3. Rated power supply time: 90 minutes
4. Charging time: 4 hours
working environment:
1. Temperature: 10~40℃
2. Humidity: 5~95%, no condensation
3. Ambient pressure: 70~106kPa
Storage environment:
1. Temperature: -20~60℃
2. Humidity: 10~95%, no condensation
3. Ambient pressure: 50~106kPa
Call the police:
Minute ventilation, airway pressure, expiratory tidal volume, respiratory rate, inspired oxygen concentration, apnea time, battery low, battery exhausted, low oxygen source pressure
Communication Interface:
USB interface, calibration interface, serial port RS232, VGA standard output data interface